						Size of Each Unit (W*D*H)(mm)W*D*H: 443*615*125(mm)
- 
						AC InputAC 220V +10%/ -20%, 50Hz;
- 
						ResolutionAD: 16bit;DA: 16bit
- 
						Input Impedance≥ 500MΩ 开机状态(关机时漏电流为100uA)
- 
						Input Active Power500w
- 
						Channel Characteristics四量程 宽动态范围 高速采样 高精度
- 
						Channel Control ModeIndependent Control
- 
						VoltageConstant Voltage Range Control Charging:0.7 V~5V; Discharging:0.7 V~5V (Clamp Line Length: 0.5m ); 
 Minimum Discharge Voltage 0.7 V (Clamp Line Length: 0.5m ); 
 Accuracy ± 0.02% of FS; 
 Stability ± 0.005% of FS; 
- 
						CurrentOutput Range per Channel Range 1: 0.1uA---150uA; Range 2:150uA---5mA; Range 3:5mA---150mA; Range 4:150mA---5A; 
 
 Accuracy ± 0.02% of FS; 
 Constant Voltage Cut-off Current 2μA;0.01mA;0.02mA; 
 Stability ± 0.005% of FS; 
- 
						PowerSingle Channel Maximum Output Power: 25W; Stability: ± 0.01% of FS; 
- 
						TimeCurrent Response Time (10%FS~90%FS) ≤100μS (10% to 90% or 90% to 10%); Step Time Range ≥10ms 
- 
						Data RecordData Recording Conditions 
 Time △t:≥10ms; Voltage △U:≥1mV; Current △I:≥100nA; 
 
 Data Acquisition Frequency 100Hz; (Continuous Charge & Discharge Mode); 
- 
						ChargeCharging Mode: Constant Current Charging, Constant Voltage Charging, Constant Current/ Constant Voltage Charging, Constant Power Charging, Constant Resistance Charging; Cut-off Conditions: Voltage, Current, Relative Time, Capacity, Energy, Power; 
- 
						DischargeDischarge Mode: Constant Current Discharge, Constant Power Discharge and Constant Resistance Discharge; Cut-off Conditions: Voltage, Current, Relative Time, Capacity; 
- 
						DCIR TestSupport DCIR Working Steps
- 
						Cycle LifeCycle Test Range: 1~65535 times; Single Cycle Steps: 254; Loop nesting: it has the function of nested loop and supports up to 4 levels of nesting; 
- 
						ProtectSoftware Protection Power down Data Protection; Function of Offline Test; The safety protection conditions can be set, and the setting specs include: upper voltage limit, lower voltage limit, upper current limit, lower current limit, upper capacity limit, Voltage Fluctuation Protection, Current Fluctuation Protection; 
- 
						Voltage and Current SamplingFour Wire Connection
- 
						DatabaseAdopting MySQL database to centralized manage testing data.
- 
						Communication Mode of Upper ComputerBased on TCP/ IP Protocol
- 
						Data Output ModeEXCEL2003,2010、TXT
- 
						Communication InterfaceNetwork Interface
- 
						Total Number of Channels8
- 
						Clamp TypeUniversal Clamps
- 
						Server Operating SystemWindows 7 64Bit
- 
						Working Temperature Range25±5℃ (Accuracy Assurance),25±20℃ (Extreme Working Temperature);
- 
						Storage Temperature Range0~60℃
- 
						Relative Humidity Range of Working Environment≤70% RH(There is no condensation of moisture.)
- 
						Relative Humidity Range of Storage Environment≤80% RH(There is no condensation of moisture.)
- 
						Clamp Picture
 
  
  
  
  
 
           
  
  
  
  
    
  
                     
  
                     
  
  
 


 
  
  
  
  
  
  
  
  
  
  
  
  언어