						Size of Each Unit (W*D*H)(mm)2U(19")443*330*84.5
- 
						AC InputAC 220V ±10% / 50Hz
- 
						Input Active Power50W
- 
						ResolutionAD:16bit;DA:16bit
- 
						Input Impedance≥1GΩ
- 
						VoltageConstant Voltage Range Control: 25mV~5V; Minimum Discharge Voltage: Upper and lower clamps can discharge to 1V, and 2m wire length can reach 1.5V; Accuracy: ± 0.05% of FS; Stability: ± 0.05% of FS; 
- 
						Current5V0.6A Output Range per Channel: 0.5mA-0.1A;0.1A-0.3A;0.3A-0.6A Accuracy: ± 0.05% of FS; Constant Voltage Cut-off Current: 0.2mA;0.6mA;1.2mA Stability: ± 0.05% of FS; 
 5V1A Output Range per Channel: 0.5mA-0.1A;0.1A-0.5A;0.5A-1A; Accuracy: ± 0.05% of FS; Constant Voltage Cut-off Current: 0.2mA;1mA;2mA; Stability: ± 0.05% of FS; 
- 
						Power5V0.6A Single Channel Maximum Output Power: 3W; Stability: ± 0.1% of FS; 
 5V1A Single Channel Maximum Output Power: 5W; Stability: ± 0.1% of FS; 
- 
						TimeCurrent Response Time (10%FS~90%FS) <1ms; Step Time Range ① ≤(365*24) h/step;② Time Format Support 00:00:00:00(h, min, s, ms) 
- 
						Data RecordData Recording Conditions Minimum Time Interval: 100ms; Minimum Voltage Interval: 10mV; 0.6A Minimum Current Interval: 0.2mA;0.6mA;1.2mA; 1A Minimum Current Interval: 0.2mA;1mA;2mA; 
 Data Acquisition Frequency 10Hz; 
- 
						ChargeCharging Mode: Constant Current Charging, Constant Voltage Charging, Constant Current/ Constant Voltage Charging, Constant Power Charging; Cut-off Conditions: Voltage, Current, Relative Time, Capacity, Energy, -△V; 
 
 
- 
						DischargeDischarge Mode: Constant Current Discharge, Constant Voltage Discharge, Constant Current/ Voltage Discharge, Constant Power Discharge and Constant Resistance Discharge; Cut-off Conditions: Voltage, Current, Relative Time, Capacity and Energy; 
 
- 
						Pulse ModeCharging: Constant Current Mode, Constant Power Mode; Discharge: Constant Current Mode, Constant Power Mode; Minimum Pulse Width: 500ms; Pulse Number: a single pulse step supports 32 different pulses; Continuous Switching between Charging and Discharging: One Pulse Step can Realize Continuous Switching from Charging to Discharging; Cut-off Conditions: Voltage, Relative time; 
 
- 
						DCIR TestSupport DCIR calculating with user defined point.
- 
						Cycle LifeCycle Test Range:1~65535 Times; Single Cycle Steps:254; Loop nesting: it has the function of nested loop and supports up to 3 levels of nesting; 
 
 
- 
						ProtectPower down Data Protection; Function of Offline Test; The safety protection conditions can be set, and the setting specs include: upper voltage limit, lower voltage limit, upper current limit, lower current limit, upper capacity limit and delay time. 
 
- 
						IP Protection DegreeProtection Degree: IP20
- 
						Channel CharacteristicsConstant current and constant voltage adopt double closed loop structure.
- 
						Channel Control ModeIndependent Control
- 
						Voltage and Current SamplingFour Wire Connection
- 
						Noise<55dB
- 
						DatabaseAdopting MySQL database to centralized manage testing data.
- 
						Communication Mode of Upper ComputerBased on TCP/ IP Protocol
- 
						Server Operating SystemWindows 7 or Windows 10
- 
						Data Output ModeEXCEL2003,2010、TXT
- 
						Server Disk Configuration500GB
- 
						Communication InterfaceNetwork Interface
- 
						Leakage Current<5μA
- 
						Total Number of Channels8
- 
						Working Temperature Range0℃~40℃(Within the range of 25±10℃, accuracy drift is 0.005% of FS /℃)
- 
						Storage Temperature Range-10℃~50℃
- 
						Relative Humidity Range of Working Environment≤70% RH(There is no condensation of moisture.)
- 
						Relative Humidity Range of Storage Environment≤80% RH(There is no condensation of moisture.)
- 
						Clamp TypeUpper and Lower Clamps
- 
						Clamp Picture
 
  
  
  
  
 
           
  
  
  
  
    
  
                     
  
                     
  
  
 


 
  
  
  
  
  
  
  
  
  
  
  
  언어